Product Detail
Model No.:
XG1.ODH-0.8
Production Capacity:
800~1000 UNITS/MONTH
Delivery Date:
15 DAYS
Function:
Pharmaceutical Companies & ...
Pipeline System:
Yes
Certificate:
CE, ISO900, 13485
Means of Transport:
BY SEA /BY AIR
Packing:
CASE
Brand Name:
YONGFENG
Structure:
Horizontal Rectangular Stru...
Touch Screen Monitoring System:
Yes
800L Pharmaceutical Ampoule Sterilizer (Mechanized Double Door), horizontal rectangular structure, touch screen monitoring system, pipeline system...
Pharmaceutical Ampoule sterilizer |
Type:XG1.ODH-0.8Mechanized door ampoule sterilizer (double door) | Made Department:Disinfection and sterilization equipment branch | Overall Dimensions:1742×1280×1920mm | Net Weight:1600kg | | |
|
USAGE |
| Product Introduction This series of sterilizers are mainly used for sterilization, leak-hunting and cleaning of ampoule, oral liquid, injection bottles and etc in pharmaceutical companies and medical institutions. Pressure steam is used as the sterilization medium with many spots for permutation of steam into air to increase temperature, which ensures a complete air exhaust and uniform temperature increase within 105℃~127℃.After sterilization, color water leak-hunting and vacuum leak-hunting can be performed on ampoule and oral liquid. These products are in full accordance with ISO9001 quality standard and GMP regulations in design, manufacture and inspection. | |
|
Specifications |
| Technical Parameter Chamber Size: 1450×610×910mm Chamber Volume: 0.8 cubic meter Max Steam Consumption: 70kg/cycle Max Color Water Consumption: 800kg/cycle Max Deionized Water Consumption: 600kg/cycle Max Compressed Air Consumption: 1.5 cubic meter Dynamic Power: :(3-AC380V/50Hz/kw): 2.0 kW | |
|
Characteristics |
| Performance characteristic ·The main body is a horizontal rectangular structure manufactured under strict quality and safety standard. ·The safe, reliable airproof door easy to be manipulated. ·The reliable, nice and compact pipeline system convenient for maintenance. ·The strong, realistic and loading trolley and specialized structure trays. ·The stable and reliable touch screen monitoring system, convenient for operation. ·Two kinds of methods of leak-hunting for ampoule and one for oral liquid available for choice. ·Extractive stainless steel cover. ·GMP validation interface is equipped for validation anywhere and anytime. | |
|
|
| | | | Configuration ·The inner shell is made of 304 stainless steel and polished. ·Slopy pneumatic valve from GEMU Germany. ·Supercharge pump from STERLING Germany. ·Germany nash_elmo Vaccum Pump. ·PLC+DCS remote control computer system from SIEMENS Germany. ·Manual double door for all. ·Monorail system ·Sterilizing trolley size: 1160x545x720mm ·quantity of sterilizing trolley:1 ·1-2ml tray size:280×265×75mm ·quantity of 1-2ml tray:72 trays ·Size of 5ml sterilizing trays:280×265×90mm ·Quantity of 5ml trays:64 trays · Size of10ml trays:280×265×105mm ·Quantity of 10ml trays:48 trays · Size of 20ml trays: 280×265×130mm ·Quantity of 20ml trays:40 trays ·Loading quantity of 1ml ampoules:48600 pcs ·Loading quantity of 2ml ampoules:43000pcs ·Loading quantity of 5ml ampoules:18400pcs ·Loading quantity of 10ml: ampoules : 10700pcs ·Loading quantity of 20ml ampoules:5760pcs |
|
|
|
| Installation requirements ·Steam: 0.3-0.5Mpa saturated ·Color water: 0.15-0.3Mpa ·Deionized water: 0.2-0.4Mpa ·Compressed air: 0.4-0.7Mpa dust removed and water, oil filtered | |
|
·First sterilizing, then leak hunting, or first leak hunting, then sterilizing for the programs, leak hunting is done by means of both vacuum and posive pressure. Two times cleaning and two adding-pressure pumping. |